Air-kerma, entrance surface

675 bytes added, 17:50, 6 June 2019
Created page with "The air kerma on the central x-ray beam axis at the point where the x-ray beam enters the patient. The contribution of backscatter radiation is included through backscatter fa..."
The air kerma on the central x-ray beam axis at the point where the x-ray beam enters the patient. The contribution of backscatter radiation is included through backscatter factor (B), thus:

<div style='text-align: center;'>
<math>K_e= B⋅K_i</math>
</div>

The unit for entrance surface air kerma is the Gy. Entrance surface air kerma Ke can be calculated from incident air kerma <math>K_i</math> using suitable backscatter factor B, or determined directly using small dosimeters (thermoluminescent or semiconductor) positioned at the representative point on the skin of the patients. (Adapted from [[ICRP Publication 117]], 2010)


''[[ICRP Glossary]] entry - May 2019''